DMP 333
Industrial
Pressure Transmitter
For High Pressure
Stainless Steel Sensor
accuracy according to IEC 60770:
standard: 0.35 % FSO
option: 0.25 / 0.1 % FSO
Nominal pressure
from 0 ... 60 bar up to 0 ... 600 bar
Output signals
2-wire: 4 ... 20 mA
3-wire: 0 … 20 mA / 0 … 10 V
others on request
Special characteristics
excellent long-term stability, also with
high dynamic pressure loads
insensitive to pressure peaks
high overpressure capability
Optional versions
IS-version
Ex ia = intrinsically safe for gases and
dusts
SIL 2 version
according to IEC 61508 / IEC 61511
customer specific versions
The pressure transmitter type DMP
333 has
been especially designed for use
in hydraulic
applications with high static and dynamic pre
s-
sure. The transmitter is characterized by an
excellent
long term stability, also under fast
changing pressure as well as positive and neg
a-
tive pressure peaks.
T
he modular concept of the device allows to
combine different stainless steel sensors and
electronic modules with a variety of electrical
and mechanical versions.Thus a diversity of
variations is created, meeting almost all r
e-
quirements in hydraulic applica
tions.
Preferred areas of use are
Plant and Machine Engineering
- machine tools
- hydraulic presses
- injection moulding machine
- handling equipment
- elevated platforms
- test benches
Mobile Hydraulics
DMP 333
Industrial Pressure Transmitter Technical Data
pressure and level
Input pressure range
Nominal pressure
gauge
1
/ abs.
[bar]
60 100 160 250 400 600
210
600
600
1000
1000
1000
420
1000
1000
1250
1250
1800
1
measurement starts with ambient pressure
Output signal / Supply
Standard
2-wire: 4 ... 20 mA / V
S
= 8 ... 32 V
DC
Option IS-protection
2-wire: 4 ... 20 mA / V
S
= 10 ... 28 V
DC
Options 3-wire
3-wire: 0 ... 20 mA / V
S
= 14 ... 30 V
DC
0 ... 10 V / V
S
= 14 ... 30 V
DC
Performance
Accuracy
2
standard: ± 0.35 % FSO
option 1: ± 0.25 % FSO
option 2: ± 0.1 % FSO
Permissible load
current 2-wire: R
max
= [(V
S
– V
S
min) / 0.02 A]
current 3-wire: R
max
= 500
voltage 3-wire: R
min
= 10 k
Influence effects
supply: 0.05 % FSO / 10 V
load: 0.05 % FSO / k
Long term stability
± 0.1 % FSO / year at reference conditions
Response time
2-wire: ≤ 10 msec
3-wire: 3 msec
2
accuracy according to IEC 60770 limit point adjustment (non-linearity, hysteresis, repeatability)
Thermal effects (Offset and Span)
Tolerance band
≤ ± 0.75 % FSO
in compensated range
0 ... 70 °C
Permissible temperatures
Permissible temperatures
medium: -40 ... 125 °C
electronics / environment: -40 ... 85 °C
storage: -40 ... 100 °C
Electrical protection
Short-circuit protection
permanent
Reverse polarity protection
no damage, but also no function
Electromagnetic compatibility
emission and immunity according to EN 61326
Mechanical stability
Vibration
10 g RMS (25 ... 2000 Hz) according to DIN EN 60068-2-6
Shock
100 g / 11 msec according to DIN EN 60068-2-27
Materials
Pressure port
stainless steel 1.4404 (316 L)
Housing
stainless steel 1.4404 (316 L)
Option compact field housing
stainless steel 1.4305 (303), cable gland brass, nickel plated others on request
Seals (media wetted)
standard: FKM
options: EPDM (for P
N
≤ 160 bar)
NBR
others on request
Diaphragm
stainless steel 1.4435 (316 L)
Media wetted parts
pressure port, seals, diaphragm
Explosion protection (only for 4 ... 20 mA / 2-wire)
Approvals
DX19-DMP 333
IBExU 10 ATEX 1068 X / IECEx IBE 12.0027X
zone 0: II 1G Ex ia IIC T4 Ga
zone 20: II 1D Ex ia IIIC T 85°C Da
Safety technical maximum values
U
i
= 28 V
DC
, I
i
= 93 mA, P
i
= 660 mW, C
i
≈ 0 nF, L
i
0 µH,
the supply connections have an inner capacity of max. 27 nF to the housing
Permissible temperatures for
environment
in zone 0: -20 ... 60 °C with p
atm
0.8 bar up to 1.1 bar
in zone 1 or higher: -20 ... 70 °C
Connecting cables (by factory)
cable capacitance: signal line/shield also signal line/signal line: 160 pF/m
cable inductance: signal line/shield also signal line/signal line: 1µH/m
DMP 333
Industrial Pressure Transmitter Technical Data
Miscellaneous
Option SIL
3
2
according to IEC 61508 / IEC 61511
Current consumption
signal output current: max. 25 mA signal output voltage: max. 7 mA
Weight
approx. 140 g
Installation position
any
4
Operational life
> 100 x 10
6
pressure cycles
CE-conformity
EMC Directive: 2004/108/EC Pressure Equipment Directive: 97/23/EC (module A)
5
ATEX Directive
94/4/EG
3
only for 4 … 20 mA / 2-wire, not in combination with the accuracy 0.1%
4
Pressure transmitters are calibrated in a vertical position with the pressure connection down.
5
This directive is only valid for devices with maximum permissible overpressure > 200 bar
Wiring diagrams
2-wire-system (current) 3-wire-system (current / voltage)
Pin configuration
Electrical connection
ISO 4400
Binder 723
(5-pin)
M12x1 / metal
(4-pin)
field housing
cable colours
(DIN 47100)
Supply +
Supply
Signal + (only for 3-wire)
1
2
3
3
4
1
1
2
3
IN +
IN
OUT +
wh (white)
bn (brown)
gn (green)
Shield ground pin 5 4
ye/gn
(yellow / green)
Electrical connections (dimensions in mm)
standard option
ISO 4400 Binder Series 723 5-pin M12x1 4-pin cable outlet with PVC cable
(IP 65) (IP 67) (IP 67) (IP 67)
6
compact field housing cable outlet,
(IP 67) cable with ventilation tube
(IP 68)
7
universal field housing stainless steel 1.4404 (316 L) with cable gland M20x1.5 (ordering code 880)
and other versions on request
6
standard: 2 m PVC cable without ventilation tube (permissible temperature: -5 ... 70 °C)
7
different cable types and lengths available, permissible temperature depends on kind of cable
p
supply +
supply
V
S
I
supply +
signal +
V
S
A/V
p
I/U
supply
DMP 333
Industrial Pressure Transmitter Technical Data
Mechanical connections (
dimensions in mm
)
standard for accuracy 0.35 / 0.25 % standard for accuracy 0.1 % ;
SIL- and SIL-IS-version
G1/2" DIN 3852 G1/2" DIN 3852
with ISO 4400 with ISO 4400
option
G1/2" EN 837 G1/4" DIN 3852 G1/4" EN 837 1/2" NPT
metric threads and other versions on request
This data sheet contains product specification, properties are not guaranteed. Subject to change without notice.
DMP333_E_151113
59
- - - - - - -
Pressure
1 3 0
1 3 1
Input [bar]
60 6 0 0 2
100 1 0 0 3
160 1 6 0 3
250 2 5 0 3
400 4 0 0 3
600
6 0 0 3
9
9 9 9 consult
Output
1
2
3
E
1S
9 consult
Accuracy
standard 0.35 % 3
option 1 0.25 % 2
option 2
0.1 %
2
1
9 consult
Electrical connection
1 0 0
2 0 0
3
T A 0
T R 0
M 1 0
stainless steel 1.4305
9 9 9 consult
Mechanical connection
1 0 0
2 0 0
3 0 0
4 0 0
N 0 0
9 9 9 consult
Seals
1
3
5
9 consult
Special version
0 0 0
9 9 9 consult
1
measurement starts with ambient pressure
2
not in combination with SIL
3
standard: 2 m PVC cable without ventilation tube (permissible temperature: -5 … 70 °C), optionally without ventilation tube
4
cable with ventilation tube (code TR0 = PVC cable), different cable types and lengths available, permissible temperature depends on kind of cable, price without cable
5
possible for nominal pressure ranges P
N
160 bar
8 5 0
4
Cable outlet
customer
EPDM
NBR
G1/4" DIN 3852
G1/4" EN 837
5
customer
1/2" NPT
FKM
customer
standard
customer
G1/2" DIN 3852
G1/2" EN 837
Male plug M12x1 (4-pin) / metal
Compact field housing
Male and female plug ISO 4400
Male plug Binder series 723 (5-pin)
Cable outlet with PVC cable
4 … 20 mA / 2-wire
customer
customer
SIL2 4 … 20 mA / 2-wire
SIL2 with Intrinsic safety
ES
0 … 20 mA / 3-wire
0 … 10 V / 3-wire
Intrinsic safety 4 … 20 mA / 2-wire
1
absolute
customer
4 … 20 mA / 2-wire
This dokument contains product specification; properties are not guaranteed. Detailed information about options are defined in the datasheet. Subject to change without notice.
Ordering code DMP 333
DMP 333
gauge
0..201