Technical data
Main specifications
Process connections See page 3
Measuring ranges Any combinations between -50 – 200 °C
Minimum span 25 °C
Output signals 4 – 20 mA or Ratiometric 10 – 90%
Electrical connections See page 4
Electrical specifications
Nom. Output signal (short-circuit protected)
4 – 20 mA
ratiometric 10 – 90%
of supply voltage
Supply voltage [U
S
] polarity
protected
10 – 30 V DC
4.75 – 8 V DC
5 V DC (Nom.)
Supply – current consumption – < 4 mA at 5 V DC
Insulation resistance
> 100 MΩ at
100 V DC
> 100 MΩ at
100 V DC
Supply voltage dependency < ± 0.05% FS/10 V
–
Current limitation 30 mA –
Output impedance – < 25 Ω
Load [RL] R
L
<(U
S
-10) / (0.02 A) ohm R
L
> 5 kohm at 5 V DC
Environmental conditions
Media temperature (max. 120 °C without extension length) -50 – 200 °C
Temperature on electronics
1
)
-40 – 85 °C
Transport temperature range -50 – 85 °C
EMC – Emmision EN 61000-6-3
EMC – Immunity EN 61000-6-2
Vibration stability
Sinusoidal 15.9 mm-pp, 5 Hz – 25 Hz –
4 g, 25 Hz – 2 kHz IEC 60068-2-6
Random 7.5 g
ms
, 5 Hz – 1 kHz IEC 600868-2-34, IEC 60068-2-36
Shock resistance
Shock 500 g/1 ms IEC 60068-2-27
Free fall IEC 60068-2-32
Enclosure (depending on electrical connections) See page 4
1
)
Temperature of the electronics depends on the media temperature, extension length, ambient temperature and air velocity.
Performance
Indicative response times
Water 0.2 m/s Air 1 m/s
ø8 mm
t
0.5
t
0.9
t
0.5
t
0.9
10 s 35 s 95 s 310 s
Accuracy
< ± 0.5% FS (typ.)
< ± 1.0% FS (max.)
Max. load protection tube
100 bar
Mechanical characteristics
Materials:
Wetted parts
Enclosure
W.no. 1.4571 (AISI 316 Ti)
W.no. 1.4404 (AISI 316 L)
Measuring insert fixed
Net weight
(Depending on design)
0.1 – 0.15 kg